Ripple Noise Meter Keisoku RM-103

Main Features:

Detailed Description

  • AC Ripple Noise and Switching Ripple can be measured independently
  • Human reading errors can be completely eliminated
  • Only one Ripple Noise Meter which conforms to the Standard of EIAJ, Electronic Industries Association of Japan
  • Only one Ripple Noise Meter which can measure faster noise up tp 100MHz
  • Automatic Ripple Ratio Selection Function: Ripple Ratio is Parameter to measure p-p Ripple Noise accurately  

A

Ripple & Noise (LF+HF, NOISE)

B

Ripple (LF+HF, RIPPLE)

C

Noise (HF, NOISE)

D

Switching Ripple (HF, RIPPLE)

E

AC Ripple (LF, Ripple)

Ripple Noise Meter Keisoku RM-103

Specifications of 884xA Series Multimeters

DC Volt.

Measurement

Range

±6.0000V

±60.000V

±500.00V

Resolution

0.1mV

1.0mV

10.0mV

Accuracy

±0.025% of rdg. ±0.025% of f.s.

Max. Applied Volt.

±500V

Measurement Time

less than 90ms (Fast Mode)/less than250ms (Slow Mode)

Ripple Noise Measurement

Ranges

300.0mVp-p

3000mVp-p

Resolution

0.1mV

1.0mV

Accuracy

±2% of rdg. ±1% of f.s.

Nominal Bandwidth

100MHz

LF Bandwidth

50Hz - 2KHz

HF Bandwidth

2KHz-

THRU Bandwidth

50Hz-100MHz

Band Pass Filter

20MHz

Ripple Ratio

0.0% - 50% (Resolution 0.5%)

Measurement Time

approx.170ms (TYP)

Allowed Input

4Vp-p

Input Condition

Impedance

DC: 1MOhm,High Frequency: 50Ohm

Input Cable

50Ohm Signal Cable RG-58/Differential Probe DP-100

Interface

GP-IB

Conform to IEEE488-1987

PASS/FAIL Judgment

Photo Coupler Output same Common (14 pin Connector)

Start Trigger

Photo Coupler Input same Common (14 pin Connector)

4CH Output

Photo Coupler Output for SC-81 Control

Input Connector

BNC Connector

Other Function

Automatic Rippl Noise Ratio Selection Function

General

Input Voltage

AC90-264V (50Hz/60Hz)

Consumption Power

20VA

Weight approx.1.7kg

Dimensions

180 (W) × 85 (H) × 300 (D)

Other Names:

RM103, RM 103

EMS supplier